Probing fails or nozzle never touches the bed — loadcell noise¶
Summary¶
If bed probing behaves as though the nozzle has already touched down while it is still obviously well clear of the sheet, the likely cause is electrical interference in the loadcell signal rather than anything mechanical. The vendor has acknowledged interference from the heater as the working theory. The community fix, which vendor support now also recommends, is a clamp-on ferrite core on the main toolhead cable close to where it enters the controller board. Several owners report this resolving the fault outright. Owners have since named further sources. Switching the LED status strip off has stopped the failures for several of them, costs nothing and is easily undone, so it is worth trying first; the bed heater wiring and the toolhead power wires are less well supported. All of these are covered below. A loadcell self-test that reports noise straight after a toolchange (possibly also one the test makes itself) may be a different matter: noise from the extruder motor, which one issue traced to the motor being left switched on after a toolchange and a second measured on 7.0.0. See below before buying hardware for that.
Error codes that lead here¶
| Code | What the printer shows |
|---|---|
36526 |
Loadcell measure failed |
36527 |
Loadcell bad configuration |
36528 |
Loadcell timeout |
All three are loadcell faults. Interference tends to show as a failed measurement or a timeout rather than a configuration error.
Detail¶
The INDX senses bed contact through a loadcell, and the signal from it shares a cable loom with the power that drives the heater. In that loom the power pair is twisted, which cancels most of its radiated noise, but the loadcell's signal pair is not — so it is comparatively exposed to whatever the heater is doing. When that noise is large enough, the firmware reads it as a contact event.
The tell-tale is how wrong the behavior is. A mechanical or offset problem makes the nozzle probe slightly too high or too low. Noise-driven false contact makes it stop while the nozzle is plainly nowhere near the sheet — a gap you can see across the room, not one you would measure with paper. If you are watching a probing pass and thinking "it hasn't even got close", this page is probably your fault.
Symptoms reported in this family:
- Probing completes with the nozzle visibly clear of the sheet
- Loadcell self-test failures
- Z homing or probing that fails only once the hotend is hot — a strong hint, because it points at the heater as the noise source
- False Z-collision errors
- Tool offset calibration that fails as a print starts but passes when run on its own from the menu. Owners describe both the heaters and the LED status strip as off during the menu run, so this pattern fits more than one source; see the section on other sources below
- Repeated bed-leveling retries, and unusually long mesh times
- A first layer that does not stick, or that drags filament up into a blob, because the machine believes the bed is higher than it is
A noisy self-test is not always interference¶
In one report
(#5468), the loadcell
test in the first calibration after a C1 was upgraded to a C1+ (Gen 2) with INDX, on
6.9.0, never sounded its beeps and rejected the press, reporting either a premature
press or a noisy signal; once the owner aborted the wizard and started it again, it
passed straight away. The owner suspected that silent mode, set on the pre-INDX
firmware, had carried over. Prusa replied that, as far as it knew, the conversion
factory-resets the printer, and that it could not reproduce the fault. A second owner,
also on a Gen 2 upgrade, describes the same fail-then-pass-on-retry outcome in
a forum thread:
the self-test kept flagging a noisy signal until they aborted it, homed the printer and
ran it again, and then it passed. Only that outcome is shared. The first report
concerned only the first run after the conversion, with the beeps missing; the second
owner's failure was not tied to a first run: it came both in the full calibration
sequence and when the test was started on its own from a cold machine, and they
describe a faint hum during the failing attempt.
Two reports in different places make the retry outcome reported, but neither
explains it. A retry costs nothing: if the self-test fails, abort and run it once more
before treating it as interference. The second owner also homed the printer before the
retry; that step comes from their report alone, provisional, though the report below
gives a reason it could matter.
A later report offers a firmware explanation that would fit both. In #5518, an owner with an eight-tool Gen 2 upgrade on 6.9.1 found the self-test failing as noisy every time after a toolchange and passing again after homing, with each of the four tools they tried. They traced it to the extruder motor, which on the INDX also works the tool lock: a toolchange leaves that motor switched on, and while it is on, the loadcell signal is noisy enough to fail the test. Switching that motor on by itself, with no toolchange, made the signal noisy; switching it off after a toolchange made it clean again without homing. Homing clears it because homing Z is itself a probe of the bed, and probing switches the extruder motor off first; homing only X and Y left the noise in place. Moving cables, switching fans and heaters, and changing temperatures neither caused it nor cleared it, which is what sets it apart from the heater interference on the rest of this page. Another owner, in the same issue, heard something from inside the head during the failing test, found that aborting the test switched the motor off, and passed on the retry. The firmware source fits the reporter's reading: in 6.9.1 the probing code switches the extruder motor off before probing, with a comment that this reduces noise on the sensor; the INDX loadcell self-test does not; and the toolchange code puts back the motor's current and position after working the lock, but not whether it was on.
That would account for the two earlier reports: an abort that switches the motor off, a homing pass before the retry, and the faint hum the second owner heard. Neither of those owners tested it, so the link is an inference. Neither says whether a tool had just been changed, either. In 6.9.1, though, the self-test picks up a tool by itself when none is held, and 6.9.0, which the first report was on, has the same step; that pickup would leave the motor on in the same way. That too is a reading of the code, not a tested result.
A separate issue, #5548, reports the same effect on 7.0.0, from the owner of another eight-tool Gen 2 machine. After an ordinary tool pickup their self-test often reports noise with nobody touching the nozzle; retrying can leave it noisy for minutes, while aborting and starting a fresh test can make it quiet. With the head standing still on a cold tool and no heater on, they recorded the loadcell with the extruder motor commanded off, then on, then off again: the noise was about three times as high while the motor was on.
Whether the same motor also upsets probing when a print starts is open, and the newest reports point partly elsewhere. The probing code switches the motor off first, yet owners in both issues have seen print-start probing fail right after a tool pickup. The first reporter saw tool offset calibration fail when a toolchange came before it, and got prints going by selecting the starting tool and homing before printing. The 7.0.0 owner's prints stop on 36130 during the loadcell probe of the offset sensor; the motor was off once nozzle cleaning had finished, but they could not record its state during the failing probe. They replaced the nozzle-cleaning macro with their own version on the USB stick, which switches the motor off before that probe, and added start G-code lines that restore the driver afterwards; they report far fewer failed starts, though not none. The macro writes the motor driver's configuration directly, nobody else has checked it, and it is not reproduced here. A third owner in that issue has had tool offset calibration trouble on their eight-tool machine since the first release; no firmware release has cured it reliably, and it still turned up on nearly every print. They switched the extruder motor off, changing nothing else, and the probing passed at the first attempt after several failures. That is a single attempt.
The owner who heard the head has a different result. A replacement head cable from Prusa
support changed nothing. Their routine of picking up the first tool from the Control
menu, running the loadcell test, aborting it once it failed and only then starting the
print worked only while the print needed no other tool before homing; once a print had
the printer pick up a second tool first, which on their machine happened when the
chamber vent had to be opened, it failed again. Switching the extruder motor off in the
start G-code (M84 E) before Z homing seemed to help at first, but homing at print
start went on failing, even with the nozzle heater switched off too and the wiring at
the back wrapped in foil tape and kept clear of the bed's power wiring. Prusa support
then advised replacing the head because the loadcell appeared defective. Before doing
so, they switched the LED status strip off: homing passed on four prints in a row with
the strip off, from the second of them with the motor and heater changes taken back out,
and mesh probing needed far fewer repeat probes. The self-test still reports noise while
the motor is on, so on that machine the two look separate: the motor upsets the
self-test, while the strip appears to be what upset print-start probing. Four prints on
one machine is a small test, provisional. The 7.0.0 owner now also switches the strip
off with M150 in the start G-code before homing and back on once probing is done,
alongside the motor change, and finds starts very reliable so far; that changes two
things at once, so it does not separate them. None of these owners has shown that the
motor causes print-start failures, and the first reporter says as much.
The effect on the self-test now rests on two separate issues, which makes it
reported. The code reading behind it is the first reporter's and this page's, not
Prusa's, and the link to print-start failures is provisional. Prusa had not answered
in either issue by the date above. 6.9.2 changes nothing that reaches an INDX beyond
filament presets, and none of the 7.0.0, 7.0.1 and 8.0.0
release notes, published on
2026-10-15, 2026-10-19 and 2026-10-22, list a change to the loadcell self-test or to
the extruder motor after a toolchange. The second issue shows the effect on 7.0.0;
nobody has yet reported on it under 8.0.0.
In practice: if the self-test reports noise after a toolchange, or when the test had to
pick up a tool first, home all axes, not only X and Y, or abort the test, and run it
again before suspecting interference or the loadcell.
Whether a noisy self-test means noisy probing is not settled. The second owner of the
two earlier reports and a third, in
a separate thread,
both linked the self-test's noise complaints to first-layer trouble and asked whether
one causes the other. The second owner has one area of the bed where the first layer
consistently fails to attach and suspects the loadcell, without a diagnosis. In that
owner's thread, two other machines had probing fail at one place, the back right of the
bed. One owner saw wires bunching under the bed there and later found the heated bed's
cable routed wrongly where it passes behind the rod, though they named other faults too
and were still sorting out leveling afterwards. On the other machine, an old bed spacer
had been refitted, and the correct spacer cured it (see the
assembly notes). A later owner has the same failure on
prints that cover the whole bed, still undiagnosed. Each is one report, provisional. A
fault that stays tied to one spot is more likely mechanical than electrical, so check
that first; that is an inference. The third,
whose self-test often complains of noise, tried to check: they ran the same mesh bed
leveling twice in a row, changing nothing except cleaning the sheet between runs, and
read back the probed values. They found the two runs close at every point, judged that
within tolerance, and asked whether that meant the noise warnings were nothing to worry
about.
TODO(verify): the spread between the two runs, as given in the loadcell noise and mesh
bed leveling thread; no band separating a healthy repeat from a faulty one has been
established here. Each of these is a single report, provisional, and none of the
three owners fitted a ferrite, so they say nothing either way about the fix below. A
repeated mesh may still be a cheap check before buying hardware, but only if it is run
the way a print probes, with the hotend hot, because the working theory on this page
is interference from the heater. The thread does not say whether the nozzle was hot
during those runs. Two matching runs with a cold nozzle do not rule interference out. This is an
inference, not a tested rule. If the explanation in #5518 holds, there is a further
reason the two can disagree: probing switches the extruder motor off and the self-test
does not, so a self-test that fails for that reason does not by itself mean probing is
noisy, although the print-start reports above leave that open. Neither owner checked
whether that was their case.
What to try¶
- Switch the LED status strip off as a free first test. Several owners' failures stopped with it off; the details and the menu option are under the next heading. It costs nothing and is undone in the same menu, so try it before buying or fitting anything.
- Fit a clamp-on ferrite core to the main toolhead cable, positioned near the controller board end. This is the fix with the most independent confirmation, and it is now what vendor support suggests. A plain clamp-on core over the outside of the cable has been enough for several machines.
- If a plain core is not enough, stubborn cases have responded to routing the cable through a higher-grade toroid for several turns instead of a single pass. TODO(verify): the specific ferrite material grade and the number of turns. Reported in the common-problems summary thread; not independently confirmed here.
- Recalibrate or factory-reset afterwards. On several machines the ferrite appeared to do nothing until the stored calibration data was discarded — by a factory reset or a full recalibration — because the printer was still working from figures it had captured while the signal was noisy. If you fit a core and nothing changes, do this before concluding the core did not help.
- Mind the placement. At least one reported position — at the controller's extension-board connector rather than on the main cable — made the problem worse. If your first placement degrades things, move it rather than abandoning the idea.
Other sources owners have found¶
A ferrite on the main toolhead cable treats one path for the noise. Owners in the
tool offset calibration thread
and in #5526 and
#5518 have since found
others. The bed heater wiring and the toolhead power wires each rest on a few owners,
most of whom changed several things at once, so they are provisional. Switching the
status strip off now has different owners behind it in both issues, and one more in the
forum thread alongside other changes, reported.
- The bed heater wiring. One owner, whose INDX failed tool offset calibration from its very first print, separated the two heater wires out of the main bundle in the wiring box at the back (from the routing they describe, apparently the bed heater's), twisted them together and fitted ferrite chokes where they meet the controller board, with chokes on the toolhead and offset sensor cables at the same time; it has worked since. Another owner, whose loadcell self-test sometimes reported noise, with sharp spikes on its bar graph, twisted the bed heater wires, fitted chokes to them and to the toolhead harness, and kept the cables apart and free of loops; after that the test settled and the failures stopped.
- The toolhead power wires. A third owner twisted the power and ground wires of the INDX toolhead harness into pairs along its whole length, which meant releasing the pins from the connector at the controller end one by one; that cleared most of their calibration failures. A pin put back in the wrong position can damage the electronics, so this is not a first step. The Detail section above says the power pair in the loom is twisted; this owner twisted the pairs themselves, implying theirs were not, while the second owner above describes theirs as already twisted, so harnesses may differ.
- The LED status strip. In #5526, an owner on 6.9.1 traced tool offset calibration failures during prints to the RGB status strip, the LED strip whose color follows the printer's state; they call it the bed LED. The loadcell noise rose whenever the strip was lit and fell back when it went dark. In that report, calibration failed inside a print, where the strip is lit, and passed when started from the menu, where it is not. Much of that owner's testing was done on firmware they had modified themselves, which this page does not rely on. The owner who twisted the toolhead power wires, on stock 6.9.2, still had an occasional failure after that; switching the strip off (the RGB Status Bar option, in the settings or under Tune > Lights during a print) stopped it, as they reported both in the forum thread and in #5526, and they later fitted ferrite beads to both the bed LED and the chamber LED cables and have had no trouble since. They have since added that they also lowered the probing temperature in their start G-code, against oozing (see oozing during probing), so their result mixes several changes. A further owner switched the LEDs off and put ferrites on every cable they could reach at the same time, and their failures stopped too. Another owner in #5526, for whom ferrite beads (placement not stated) did not help, finds that switching the strip off seems to work every time; and the owner in #5518 described above had their print-start homing failures stop with it off. Whether ferrites on the strip's own cable help rests on one owner, whose result mixes several changes, while switching the strip off has not failed anyone who reported it. It costs nothing and is easily undone, which makes it a cheap test before opening the wiring.
The pattern of failing in a print and passing from the menu fits both the heaters and
the strip, since owners describe both as off during the menu run; the owner who twisted
the power wires first read it as heater interference. The owner in #5518 is the first
outside the strip owner's own testing to tell the strip and the nozzle heater apart on
one machine: homing at print start still failed with the nozzle heater off, and passed
with the strip off and the heater working as normal. The report mentions switching off
only the nozzle heater, so it does not separate the strip from the bed heater. That is
one machine over three or four prints, provisional.
How to tell interference from a genuinely bad loadcell
A ferrite core treats electrical interference. It will not fix a failed loadcell, and the two present almost identically. One owner separated them cleanly: their probing fault followed a replacement toolhead across a swap while the original head homed perfectly every time, on a recent controller board. Interference is a property of the machine and its wiring; a fault that travels with the toolhead is in the toolhead. See diagonal banding, where that swap is described. If your board is recent and the fault moves with the head, go to the vendor rather than buying ferrites.
The reverse also happens. In
#5518, Prusa support
advised an owner to replace the head because the loadcell appeared defective; before
the swap, switching the status strip off stopped their print-start homing failures
over the four prints since (provisional, described above). Their self-test still
reports noise while the extruder motor is on, which fits the separate motor effect
described above; that alone is no sign of a bad loadcell (an inference). Before
accepting a head swap, try switching the strip off and aborting and rerunning the
self-test, both of which cost nothing.
A report from another thread ended in a hardware fix rather than interference. An
owner on 6.6.0 and 6.6.1 whose loadcell
test kept warning of unstable readings, although a press on the nozzle still
registered once the warning was dismissed, and whose prints stopped on a 36130 tool
offset error, eventually
traced the problem to a faulty part
they call the tool holder, which they took up with the manufacturer under warranty;
everything has worked since. The post does not come back to the loadcell warning;
that it cleared too is implied by "everything", not stated. The post mentions no
ferrite. It does not make clear
which assembly they mean, so that one is provisional. The same case appears under
tool offset calibration fails.
A third cause: homing that fails after the calibration passed
Firmware 6.9.1, released as stable on 2026-09-25, lists a homing fix for
printers that passed their first homing calibration but failed when homing again
later. The change Prusa describes is to motor currents; a Prusa developer later
explained in #5520
that it was meant to stop belts slipping during homing on some printers, where the
motor holding position could not resist the one doing the bumping. The
release notes
name no axis, but the only motor-current change in the firmware repository
between the beta and the stable
(compare)
is to the diagonal XY homing, not to the loadcell's Z probing. The notes also say
nothing about the loadcell or the heater, so this is a separate fault from the one
on this page, not a fix for it. If your homing failures match that description
rather than the pattern above — failing only once hot, the nozzle stopping visibly
short of the sheet — update the firmware before fitting a ferrite, but read the
next paragraph first. This rests on the vendor's release notes, its firmware
repository and that explanation alone, provisional: no owner has yet reported the
fix curing their homing failures.
Owners report what looks like a side effect of the same change, in
#5520, in a
forum thread
and in another:
since 6.9.1, homing shows an Endstop early trigger message several times and
takes longer. For some it still finishes and prints run normally; for others the
homing calibration fails, which stops the setup. Calibrating homing again after the
update, Prusa's first suggestion, did not help; going back to the 6.9.1 beta or to
6.9.0 made it stop for most of the owners who reported trying it. One still saw it
on 6.9.0, though less often, and another still had slow homing at times during
prints on the beta. Several ran the
homing calibration again after going back. Prusa says it did not see this on the
printers it tested the change on, has said it is looking into the reports, and has
asked owners who see it to stay on the earlier firmware in the meantime; the
comment names 6.9.1, while the owners had gone back to the 6.9.1 beta or to 6.9.0.
Prusa has since asked affected owners for photos of their X and Y motor labels and
a homing log, to narrow it down to particular motor batches.
Going back to either also gives up the gantry squareness wizard, which the
comparison above shows arriving after the beta.
Several owners report the same on 7.0.0 and 7.0.1 and one on 8.0.0, and none of the
7.0.0, 7.0.1 and 8.0.0
release notes list a change to
homing currents. In the
6.9.1 source
that message marks an X/Y homing bump that stopped short and is being repeated;
the measurement gives up only if the repeats run out. It comes from the X/Y homing,
not the loadcell, and is no sign of interference. Several owners, in the issue and
in two separate forum threads, make the message and the effect of going back
reported; that the lower current causes it is an inference.
If none of that helps, particularly if the failure happens only with the heater on and you are on an early board revision, the path is hardware replacement through the vendor. One owner reported success wrapping the wiring at the controller connector in grounded shielding foil instead, which is consistent with the same root cause.
TODO(verify): the raw loadcell value ranges that distinguish a healthy machine from an affected one. The summary thread quotes an idle-value band for each, and those numbers would make this page far more diagnostic — but they need checking against firmware before being published, because a reader will use them to decide whether their machine is faulty.
This is a mitigation, not a cure
The vendor has described the ferrite as a stopgap rather than a fix, and a firmware-side improvement to loadcell handling is reportedly in progress. Neither the 6.9.0 nor the 6.9.1 release notes describe any change to how the loadcell signal is filtered or judged. The 6.9.1-beta notes do give tool offset calibration more Z probe attempts. The stable notes do not restate it, but the stable 6.9.1 is built on that beta: in the firmware repository it adds 15 commits to the beta tag, none of them in the tool offset calibration code. That comes from the release history, not from the notes. More retries give a noisy tap more chances to pass; they do nothing about the noise. 6.9.2, released on 2026-10-07, adds the PVA and BVOH filament presets that 6.9.1 announced and left out, and nothing else that reaches an INDX (compare), so it changes nothing described on this page. 7.0.0, published on 2026-10-15, lists nothing on how the loadcell signal is filtered or judged, nor on the self-test, the extruder motor after a toolchange or the status strip. Prusa had not published its source on GitHub by the date above, so what it changes is known only from those notes and from owners: the extruder motor effect has been measured on it. 7.0.1, published on 2026-10-19, lists a single fix, for a freeze at startup, and so names none of those either. 8.0.0, published on 2026-10-22 and likewise without source on GitHub, is the same: its notes name none of those, nor homing currents. If you are reading this well after the date above, check whether a newer firmware has addressed it before adding hardware.
Verification¶
reported — independently described in more than one thread by different owners.
In Nozzle not touching bed during probing an owner reports probing running with the nozzle far above the sheet on a machine that had passed all its setup calibrations; after fitting a ferrite core to the main cable they confirm probing began working correctly. A second owner in the same thread reports one probe-too-high incident and fitted a core pre-emptively without regression. The mechanism, the vendor's acknowledgment, and the toroid variant come from the common problems summary, which is a condensation of a now-offline community knowledge base.
Where the sources are weaker: the controlled A/B test described in the summary
(failing without a core, working with one, failing again on removal) is reported
second-hand there and is not separately visible in the forum corpus. The loadcell
value bands and ferrite specifications are single-source and withheld above. The
self-test that passes on a retry now has two reports in different places, and Prusa
could not reproduce the first. The extruder motor's effect on the self-test is now in
two separate issues, one of them on 7.0.0, with a further owner in the first confirming
the abort-and-retry part; the firmware source is consistent with it, but Prusa has not
confirmed it, neither earlier owner tested it, and its link to print-start failures is
unproven: one owner's print-start failures stopped with the status strip off rather than
with the motor off. Homing before the retry, the
repeated-mesh check, the back-right probing cases, the warranty claim, the advised head
swap and the tool-detection case under Related are one report each.
None of those owners fitted a ferrite, so they neither strengthen nor weaken the case
for it. The bed heater wiring and the toolhead power wires come from one forum thread,
with the owners mostly changing several things at once. The status strip comes from two
issues and that same thread; the first owner's testing was partly on modified firmware,
but switching the strip off stopped the failures for different owners in both issues,
and in the forum thread alongside other changes, which makes that part reported.
Whether ferrites on the strip's own cable help rests on one owner, whose result mixes
several changes. The early-trigger homing message is reported by several owners across
one issue and two forum threads. The homing fix rests on the vendor's release notes, its
firmware repository and a developer's explanation, not on owner reports.
Related¶
- Tool offset calibration fails — separate sensor, separate fault, often confused with this one
- Oozing spoils bed probing and tool calibration — a completely different cause with an overlapping symptom, worth ruling out. If material is accumulating on the nozzle before contact, that is the other one.
- Tool offset calibration fails: bed not aligned in Z — a mechanical fault that was taken for loadcell noise on one machine; a ferrite core did nothing for it
- Who to contact — if it comes to a parts request: diagnosis from Prusa first, then the hardware from Bondtech.
- Assembly notes — if the loadcell test has been unstable since the machine was built and never worked, treat it as a build question before an interference one.
- Phantom tools, "tool not detected" and park failures
— how the head detects a fitted tool. A detection fault can also stop the loadcell
test without being a loadcell fault: in
one report
on 6.9.1, the test step picked a tool and drove it into the other docks, and the
owner found the printer never registered a fitted tool (
provisional). A check of the head's wiring found nothing; Prusa then replaced the toolhead, and with the new one the owner reports everything working as expected. The fault was in the head, and nothing in the report points at interference.